Equipements & Technologies

Equipements & Technologies

Equipements & Technologies de la plateforme ARPEGE

Mithras LB940

Responsable technique : Claire VOL

Applications :

  • interactions protéines-protéines,
  • production de seconds messagers
MODULE/FUNCTION DESCRIPTION
Detection Modes BRET
Measurement Modes Top and bottom reading,
Endpoint and Kinetic measurements,
Sequential Multi Excitation measurements,
Sequential Multi Emission measurements,
Simultaneous Dual Emission measurements,
Real-time ratiometric measurements,
Well Scanning
Microplate Formats Up to 384-well plates
Light Sources Halogen lamp
Detectors Low-noise photomultiplier with single photon counting technology
Excitation filters 355 nm, 485 nm
Emission filters 410 nm, 460 nm, 485 nm,
515 nm, 530 nm, 535 nm.
Sensitivity Fluorescence
<20 fmol FITC,
Luminescence
<20 amol ATP
Crosstalk Low crosstalk due to crosstalk reduction methods :
5×10-6
Temperature control +5°C above room temperature to 45°C
Reagent Injection Up to 4 injectors, JET injection technology,
Variable volumes : 10 – 100 µL,
Injection into microplates with up to 384 wells.
Shaking 3 modes : linear, orbital, double-orbital,
Variable amplitude and speed
Operation modes
Integral measurement 0.1 – 600 secondes
Kinetics measurement Total length up to 24h
Repeated measurement Total length up to 24h
Plate repeats Up to 999
Scanning Up to 100 single data points
Delay Delay