Technical specifications:
MODULE/FUNCTION | DESCRIPTION |
---|---|
Detection method | Fluorescence |
Plate formats | 96 wells |
Light source | Xenon Flash Lamp (1 joule/flash) |
Detectors | 2 photomultiplier tubes |
Exitation | 485 nm |
Emission | 538 nm |
Flex reading | Fluorescence + fluidics |
Kinetic reading | Fluorescence + fluidics |
Fluidics (96-Well Plates Only) | 8-channel |
Max. volume | 200 μL |
Precision @ 50 μL | 2% CV |
Precision @ 5 μL | 8% CV |
Dispense max. rate | 208 μL/sec. |
Sensitivity | < 5 pM fluorescein in 96 wells, < 20 pM in 384 wells |
Typical Read Times With 3 flashes/well in fluorescence mode | 96 wells : 17 seconds |
Temp. control* | 2°C above ambient to 45°C |
Temp. uniformity* | < 1°C at 37°C set point |
Temp. accuracy* | ±1°C at 37°C set point |
*Temperature is regulated in the Read Chamber | |
Shaker time | 0 to 999 seconds |