Technical specifications:

MODULE/FUNCTIONDESCRIPTION
Detection ModesFluorescence, TR-FRET/HTRF®, AlphaScreen®
Measurement ModesTop and bottom reading,
Endpoint and Kinetic measurements,
Sequential Multi Excitation measurements,
Sequential Multi Emission measurements,
Simultaneous Dual Emission measurements,
Real-time ratiometric measurements,
Well Scanning.
Microplate FormatsUp to 384-well plates, user-definable
Light SourcesHigh energy xenon flash lamp,
Nitrogen Laser (337 nm),
Solid state Laser for AlphaScreen®.
DetectorsFour photomultiplier tubes, optimized for different detection modes
Z-AdjustmentAutomatic focal height adjustment (0.1 mm resolution)
Excitation337 nm, 485 nm, 640 nm, 680 nm
Emission490 nm, 520 nm, 550 nm, 570 nm,
620 nm, 665 nm, 680 nm.
DecayTR-FRET decay measurement in real time with 2 µsec resolution
SensitivityTRF
25 fmol/L europium (0.5 amol/well) (384),
100 fmol/L europium (0.8 amol/well) (1536).
HTRF®
Reader Control Kit (Eu) after 18h (384),
Delta F > 1100 % (High Calibrator),
Delta F > 25 % (Low Calibrator).
AlphaScreen®
< 100 amol / well (384)
Read Times1 flash
14 s (384), 27 s (1536)
10 flashes
38 s (384), 1 min 52 s (1536)
50 flashes
1 min 29 s (384), 5 min 18 s (1536)
Incubation+5°C above ambient up to 45°C
Reagent InjectionUp to 2 built-in reagent injectors
Injection at measurement position (6 to 384-well)
Individual injection volumes for each well (3 to 350 µL)
Variable injection speed up to 420 µL / s
Up to four injection events per well
Reagent back flushing
ShakingLinear, orbital, and double-orbital with user-definable time and speed